Spectrophotometer CM-3600A

Mã SP: CM-3600A

The CM-3600a is a highly accurate, repeatable spectrophotometer that measures color quickly in any application. Oriented horizontally and made for benchtop use, the instrument has diameters that are interchangeable and can be used in a wide variety of measuring tasks.

Document:
Share:

 

Konica Minolta designed the CM-3600A spectrophotometer with high-precision capabilities and great reliability. Its versatility and ease of use makes it a good choice in many measuring applications. Because it’s so easy to operate, it’s consistently among the most sought after benchtop spectrophotometers on the market.

The CM-3600A spectrophotometer is a compact yet advanced instrument that features a side port in horizontal format and was built to measure a wide variety of objects’ reflectance. It also measures the transmittance of translucent or transparent liquids and solids – all this while taking up a minimum amount of space.

Numerical Gloss Control Provides Simultaneous SCI/SCE Measurements

The CM-3600A utilizes Konica Minolta’s patented Numerical Gloss Control system, eliminating the need for a mechanical gloss trap. By firing two xenon lamps in quick succession, the system can provide virtually simultaneous SCI (specular component included)/SCE (specular component excluded) measurements and also enable the calculation of 8° gloss.

Numerical UV Control for Accurate Measurements of Fluorescent Materials

Accurate measurement of materials such as paper or cloth treated with fluorescent whitening agents (FWA) requires precise control of the UV component and its effects. The patented Numerical UV Control method used by the CM-3600A can provide such control by using proprietary calculations to combine the results from flashes of two xenon lamps: one with full UV energy, the other with UV energy removed by a UV-cutoff filter (either 400 nm or 420 nm). This method eliminates the need for mechanical filter positioning and repeated filter position adjustment, and enables UV adjustment by Whiteness Index, Tint, Brightness, or UV profile.

Reflectance and Transmittance in One Instrument

The CM-3600A can measure both the reflectance of opaque objects and the transmittance of transparent or translucent solid materials such as plastics. (With accessories, the CM-3600A can even measure the transmittance of liquids.)

For reflectance measurements, the CM-3600A employ di:8°, de:8° (diffuse illumination, 8° viewing) geometry which conforms to ISO, CIE, ASTM, DIN, and JIS standards. For transmittance measurements, the CM-3600A employ d:0° (diffuse illumination, 0° viewing) geometry which conforms to ISO, CIE, ASTM, and DIN standards.

Bright, Clear Sample Viewer for Accurate Sample Positioning

Opening the sample viewer provides a clear, illuminated view of the sample at the measurement port. On the CM-3600A, this view is reflected in a mirror on the sample viewer cover so that you can view it easily even while seated.

Three Measurement Areas For Diverse Needs

Masks for the CM-3600A’s three measurement areas (SAV: Ø4 mm, MAV: Ø8 mm, and LAV: Ø25.4 mm) are included as standard accessories, so you can select the one that fits your measurement requirements.

Full Data Compatibility with CM-3600d and CM-3610d

Since the CM-3600A utilize the same design and innovative technology as their predecessors CM-3600d and full data compatibility is assured.

USB Data Communication

Data communication between the CM-3600A is performed via USB to offer higher speed.

RoHS Compliance

The CM-3600A is in full conformance with the RoHS (Restrictions on Hazardous Substances Directive) to ensure that they are environmentally friendly.

Small footprint

Despite having a 152-mm integrating sphere and being packed full of advanced functionality, the CM-3600A is extremely compact, with a footprint about the same as a sheet of B4 paper or an average laptop computer. It can fit easily in a limited space while still providing the versatility and high accuracy of larger models.

Measures both reflectance and transmittance, including transmittance of liquids

Reflectance

The CM-3600A can measure reflectance using di:8°, de:8° (diffuse illumination, 8° viewing) geometry which conforms to ISO, CIE, ASTM, DIN, and JIS standards

Transmittance

The CM-3600a can measure the transmittance of both solid and liquid materials using d:0° (diffuse illumination, 0° viewing) geometry which conforms to ISO, CIE, ASTM, and DIN standards

Improved sample holder

The sample holder opens approximately 90° for easy positioning of samples. It also has a new “soft-close” mechanism that prevents it from slamming shut and damaging samples or pinching fingers.

Bright, clear sample viewer can be used even while seated.

The sample viewer lets the sample be seen clearly at the measuring port for accurate positioning. When the sample viewer cover is opened, a high-brightness LED illuminates the sample to provide a clear image, and the cover is equipped with a mirror so that the image can be seen even from a seated position.

Specifications

Principal Specifications CM-3600A

Model Spectrophotometer CM-3600A
Illumination/observation system Reflectance; di:8°, de:8° (diffused illumination, 8-degree viewing), equipped with simultaneous measurement of SCI (specular component included) / SCE (specular component excluded); Conforms to CIE No.15, lSO7724/1, ASTM E-1164, DIN 5033 Teil7 and JIS Z8722 condition C standard.
Transmittance : di:0°, de:0° (diffused illumination, 0-degree viewing); Conforms to CIE No.15, ASTM E-1164 and DIN 5033 Teil7 standard.
Light-receiving element Silicon photodiode array (dual 40 elements)
Spectral separation device Diffraction grating
Wavelength range 360 to 740 nm
Wavelength pitch 10 nm
Half bandwidth Approx. 10 nm
Reflectance range 0 to 200%; Resolution: 0.01%
Sphere size Ø152 mm
Light source Four pulsed xenon lamps
Minimum interval between measurements Normal SCI/ SCE measurement: 4 sec.
UV-cut / UV-adjusted measurement: 5 sec.
Transmittance measurement: 3 sec.
Measurement/illumination areas Reflectance: Changeable between LAV, MAV, and SAV
LAV: Ø25.4 mm / Ø30 mm
MAV: Ø8 mm / Ø11 mm
SAV: Ø4 mm / Ø7 mm
Transmittance: Approx. Ø17 mm
Repeatability When white calibration plate is measured 30 times at 10-sec. intervals after white calibration has been performed: 
Spectral reflectance: Standard deviation within 0.1%
Colorimetric values: Standard deviation within ΔE*ab 0.02
Inter instrument agreement Mean ΔE*ab 0.15 (LAV/SCI) Average for 12 BCRA Series II color tiles compared to values measured with master body
UV adjustment Instantaneous numerical adjustment
UV cut filters 400 nm cutoff and 420 nm cutoff
Transmittance chamber Width: 133 mm, Depth: Approx. 50 mm; Measurement diameter: Approx. 17 mm
Transmission sample holder (optional accessory): Sample holder for both plate-shaped and liquid samples (removable); Liquid samples not measurable with CM-3610A.
Interface USB 1.1
Power AC 100 to 240V 50/60Hz (using included AC adapter)
Operating temperature/humidity range 13 to 33°C, relative humidity 80% or less (at 35°C) with no condensation
Storage temperature/humidity range 0 to 40°C, relative humidity 80% or less (at 35°C) with no condensation
Size (W × H × D) 244 × 205 × 378 mm
Weight 11.5 kg
Standard accessories White Calibration Plate, Target Mask (Ø4 mm), Target Mask (Ø8 mm), Target Mask (Ø25.4 mm), Zero Calibration Box, AC Adapter, Dust Cover, Accessory Case, USB Cable (3 m)
Optional accessories SpectraMagic™ NX, Transmittance Specimen Holder, Cell (2mm)/(10mm)/(20mm), Plastic Cell (2mm)/(10mm)/(20mm), Transmittance Zero Calibration Plate, USB Cable (5 m)

Specifications are subject to change without prior notice.

Related Products

CERTIFICATE OF KONICA MINOLTA AUTHORIZATION
CERTIFICATE OF KONICA MINOLTA AUTHORIZATION
Vui lòng liên hệ: 0906 362 998 (A. Kỳ) 0938 539 455(A.Thọ)
Uni Gloss 60plus Gloss Meter
Uni Gloss 60plus Gloss Meter

Measured gloss condition of object surfaces.

Vui lòng liên hệ: 0906 362 998 (A. Kỳ) 0938 539 455(A.Thọ)
Chlorophyll Meter SPAD-502Plus
Chlorophyll Meter SPAD-502Plus

The SPAD-502Plus enables quick, easy measurement of the chlorophyll content of plant leaves without damaging the leaf. Chlorophyll content is one indicator of plant health, and can be used to optimize the timing and quantity of applying additional fertilizer to provide larger crop yields of higher quality with lower environmental load.

Vui lòng liên hệ: 0906 362 998 (A. Kỳ) 0938 539 455(A.Thọ)
Spectrodensitometer FD-7/FD5
Spectrodensitometer FD-7/FD5

Even on substrates with fluorescent whitening agents, these next-generation measurement tools streamline color adjustment in printing,. The FD-7 can not only measure the evaluation light source, but then calculate data under that source.

Vui lòng liên hệ: 0906 362 998 (A. Kỳ) 0938 539 455(A.Thọ)
Spectrophotometer CM-3700A
Spectrophotometer CM-3700A

If you're looking for a highly accurate measurement instrument that uses the latest technology, then the CM-3700A bench-top spectrophotometer is just what you need, and it exceeds most industry color requirements. With a wide range of applications, the CM-3700A obtains highly-accurate reflectance and transmittance measurements.

Vui lòng liên hệ: 0906 362 998 (A. Kỳ) 0938 539 455(A.Thọ)
Spectrophotometer CM-5
Spectrophotometer CM-5

The CM-5 is a very versatile bench-top spectrophotometer to measure the colour of solid, pasty granular and liquid samples either in reflectance or transmittance. The top-port concept, offering various aperture sizes, allows measurement of almost any kind of samples. The CM-5 is thus the perfect all-rounder laboratory instrument for applications such as Flavours & Fragrances, Foods, Chemicals, Pharmaceuticals and Plastics. 

 